  Measurement system repeatability and reproducibility analysis report one-to-one mean range method
  Approve the old term:
   Preparation of the old period:
  Part Name:   Transformer TV31D- 2 mA / 2 mA 2 mA / 2 mA 2mA//2mA2 \mathrm{~mA} / 2 \mathrm{~mA}   Name of gage:   Miniature transformer error checker   Date: 2025.6.19
  Part No.: 0   Gage Number: YXCSY-CT/PCT-05   Measure Person A   Soft into Sha
  Production Batch: 0   Measurement characteristics:   ratio difference B   Zhou Qingchun
  Dimensions: -0.2 0.2   Characteristic Units: \% C   Yang Daoling
零件名称: 互感器TV31D- 2mA//2mA 量具名称: 微型互感器误差校验仪 日期: 2025.6.19 零件编号: 0 量具编号: YXCSY-CT/PCT-05 测量人 A 柔成莎 生产批次: 0 测量特性: 比差 B 周青春 尺寸规格: -0.2 0.2 特性单位: \% C 杨道玲| 零件名称: | 互感器TV31D- $2 \mathrm{~mA} / 2 \mathrm{~mA}$ | | 量具名称: | 微型互感器误差校验仪 | 日期: | 2025.6.19 | | :--- | :--- | :--- | :--- | :--- | :--- | :--- | | 零件编号: | 0 | | 量具编号: | YXCSY-CT/PCT-05 | 测量人 A | 柔成莎 | | 生产批次: | 0 | | 测量特性: | 比差 | B | 周青春 | | 尺寸规格: | -0.2 | 0.2 | 特性单位: | \% | C | 杨道玲 |
  Results:
   See: Page 2: Measurement System Repeatability and Reproducibility Data Sheet
   Page 3: Data analysis report on the repeatability and reproducibility of the measurement system
   Page 4: Mean Range Plot
   Analysis Results:
  5.15 standard deviation   % of study variability
  repeatability 0.0006 3.0 % 3.0 % 3.0%3.0 \%
  reproducibility 0.0005 2.5 % 2.5 % 2.5%2.5 \%
  Gage R\&R 0.0008 4.1 % 4.1 % 4.1%4.1 \%
  Between parts 0.0197 100.0 % 100.0 % 100.0%100.0 \%
5.15标准偏差 占研究变差百分比 重复性 0.0006 3.0% 再现性 0.0005 2.5% 量具R\&R 0.0008 4.1% 零件间 0.0197 100.0%| | 5.15标准偏差 | 占研究变差百分比 | | :---: | :---: | :---: | | 重复性 | 0.0006 | $3.0 \%$ | | 再现性 | 0.0005 | $2.5 \%$ | | 量具R\&R | 0.0008 | $4.1 \%$ | | 零件间 | 0.0197 | $100.0 \%$ |
  Concentrate:
Study variability = 0.0197 Number of Data Grading (NDC) = 35
  Result:
*Based on the NDC analysis of the discriminatory classification number, the measurement system is judged to have reliable resolution.
*Given that the sampled samples can be representative of process variation, the above calculations are based on the total variation of the study.
  Criteria for Judgment   conclusion
R & R % < 10 % R & R % < 10 % R&R% < 10%\mathrm{R} \& \mathrm{R} \%<10 \%   The Interior Measurement System is acceptable. ***\star
10 % < R & R % < 30 % 10 % < R & R % < 30 % 10% < R&R% < 30%10 \%<\mathrm{R} \& \mathrm{R} \%<30 \%
Depending on the importance of the application/cost of the installation/maintenance elements, it may be acceptable.
基于应用的重要性/则里装置成本/维修素,可能是可接受的。| 基于应用的重要性/则里装置成本/维修素,可能是可接受的。 | | :--- |
R & R % > 30 % R & R % > 30 % R&R% > 30%\mathrm{R} \& \mathrm{R} \%>30 \%
Unacceptable, all efforts should be made to improve the child-testing system.
不可接受的一一应该作出各种努力来改 进测童系统。| 不可接受的一一应该作出各种努力来改 | | :--- | | 进测童系统。 |
判断准则 结论 R&R% < 10% 测里系统是可接受的。 *** 10% < R&R% < 30% "基于应用的重要性/则里装置成本/维修素,可能是可接受的。" R&R% > 30% "不可接受的一一应该作出各种努力来改 进测童系统。" | 判断准则 | | 结论 | | :---: | :--- | :---: | | $\mathrm{R} \& \mathrm{R} \%<10 \%$ | 测里系统是可接受的。 | $\star$ | | $10 \%<\mathrm{R} \& \mathrm{R} \%<30 \%$ | 基于应用的重要性/则里装置成本/维修素,可能是可接受的。 | | | $\mathrm{R} \& \mathrm{R} \%>30 \%$ | 不可接受的一一应该作出各种努力来改 <br> 进测童系统。 | |
  Follow-up:
  Measurement System Repeatability and Reproducibility Analysis Report - Mean Range Method
  Approve the old term:
   Preparation of the old period:
  Part Name:   Instrument transformer TV 31 D 2 mA / 2 mA 31 D 2 mA / 2 mA 31D-2mA//2mA31 \mathrm{D}-2 \mathrm{~mA} / 2 \mathrm{~mA}   Name of gage:   Miniature transformer error checker   Date: 2025.6.19
  Part No.: 0   Gage Number: YXCSY-CT/PCT-05   Testers A   Soft into Sha
  Production Batch: 0   Measurement characteristics:   Angular difference B   Zhou Qingchun
  Dimensions: -20 20   Characteristic Units: C   Yang Daoling
零件名称: 互感器TV 31D-2mA//2mA 量具名称: 微型互感器误差校验仪 日期: 2025.6.19 零件编号: 0 量具编号: YXCSY-CT/PCT-05 测里人 A 柔成莎 生产批次: 0 测量特性: 角差 B 周青春 尺寸规格: -20 ~ 20 特性单位: , C 杨道玲| 零件名称: | | 互感器TV | | $31 \mathrm{D}-2 \mathrm{~mA} / 2 \mathrm{~mA}$ | 量具名称: | 微型互感器误差校验仪 | 日期: | | 2025.6.19 | | :--- | :--- | :--- | :--- | :--- | :--- | :--- | :--- | :--- | :--- | | 零件编号: | 0 | | | | 量具编号: | YXCSY-CT/PCT-05 | 测里人 | A | 柔成莎 | | 生产批次: | 0 | | | | 测量特性: | 角差 | | B | 周青春 | | 尺寸规格: | | -20 | ~ | 20 | 特性单位: | , | | C | 杨道玲 |
  Measurement Results:
   See: Page 2: Measurement System Repeatability and Reproducibility Data Sheet
   Page 3: Data analysis report on the repeatability and reproducibility of the measurement system
   Page 4: Mean Range Plot
   Analysis Results:
  5.15 standard deviation   % of study variability
  Heavy summer sex 0.06 2.8 % 2.8 % 2.8%2.8 \%
  reproducibility 0 0.0 % 0.0 % 0.0%0.0 \%
  Gage R\&R 0.06 2.8 % 2.8 % 2.8%2.8 \%
  Room 2.108 100.0 % 100.0 % 100.0%100.0 \%
5.15 标准偏差 占研究变差百分比 重夏性 0.06 2.8% 再现性 0 0.0% 量具R\&R 0.06 2.8% 雵件间 2.108 100.0%| | 5.15 标准偏差 | 占研究变差百分比 | | :---: | :---: | :---: | | 重夏性 | 0.06 | $2.8 \%$ | | 再现性 | 0 | $0.0 \%$ | | 量具R\&R | 0.06 | $2.8 \%$ | | 雵件间 | 2.108 | $100.0 \%$ |
  Concentrate:
Study variation = 2.109 2.109 2.109quad2.109 \quad Number of Data Grading (NDC) = 50
   Result:
*Based on the NDC analysis of the discriminatory classification number, the measurement system is judged to have reliable resolution.
*Given that the sampled samples can be representative of process variation, the above calculations are based on the total variation of the study.
  Criteria for Judgment   conclusion
R\&R\%<10\%   The measurement system is acceptable.
  10\%<R\&R\%<30\% May be acceptable based on factors such as the importance of the application/the cost of the device/the cost of repairs.
R\&R\%>30\%   Unacceptable – Efforts should be made to improve the measurement system.
判断准则 结论 R\&R\%<10\% 测量系统是可接受的。 * 10\%<R\&R\%<30\% 基于应用的重要性/则量装置成本/维修成本等因素,可能是可接受的。 R\&R\%>30\% 不可接受的——应该作出各种努力来改进测量系统。 | 判断准则 | | 结论 | | :--- | :--- | :--- | | R\&R\%<10\% | 测量系统是可接受的。 | * | | 10\%<R\&R\%<30\% | 基于应用的重要性/则量装置成本/维修成本等因素,可能是可接受的。 | | | R\&R\%>30\% | 不可接受的——应该作出各种努力来改进测量系统。 | |
  Follow-up:
  Measurement System Repeatability and Reproducibility Analysis Report - Mean Range Method
  Approve the old term:
   Preparation of the old period:
  Part Name:   TA31B11-5A/50mA   Name of lining:   Vernier calipers   Date: 2025.6.22
  Part No.: 0   Lining number: YXYBKC-11   Tester A
  Production Batch: 0   Polarization characteristics:   Outside diameter   Zhou Qingchun
  Dimensions: 3 4 4 ∼4\sim 4   Characteristic Units: mm B
  Yang Daoling
C   Sang Chengsha
零件名称: TA31B11-5A/50mA 里具名称: 游标卡尺 日期: 2025.6.22 零件编号: 0 里具编号: YXYBKC-11 测里人 A 生产批次: 0 测里特性: 外径 周青春 尺寸规格: 3 ∼4 特性单位: mm B 杨道玲 C 桑成莎| 零件名称: | TA31B11-5A/50mA | 里具名称: | 游标卡尺 | 日期: | 2025.6.22 | | :--- | :--- | :--- | :--- | :--- | :--- | | 零件编号: | 0 | | 里具编号: | YXYBKC-11 | 测里人 A | | 生产批次: | 0 | | 测里特性: | 外径 | 周青春 | | 尺寸规格: | 3 | $\sim 4$ | 特性单位: | mm | B | | | | | 杨道玲 | | | | | | | | C | 桑成莎 |
  Measurement Results:
   See: Page 2: Measurement System Repeatability and Reproducibility Data Sheet
   Page 3: Data analysis report on the repeatability and reproducibility of the measurement system
   Page 4: Mean Range Plot
   Analysis Results:
  5.15 standard deviation   % of study variability
  Heavy summer sex 0.0053 3.6 % 3.6 % 3.6%3.6 \%
  reproducibility 0.0125 8.6 % 8.6 % 8.6%8.6 \%
  R&R 0.0136 9.3 % 9.3 % 9.3%9.3 \%
  Between parts 0.1451 99.6 % 99.6 % 99.6%99.6 \%
5.15标准偏差 占研究变差百分比 重夏性 0.0053 3.6% 再现性 0.0125 8.6% 里具R\&R 0.0136 9.3% 零件间 0.1451 99.6%| | 5.15标准偏差 | 占研究变差百分比 | | :---: | :---: | :---: | | 重夏性 | 0.0053 | $3.6 \%$ | | 再现性 | 0.0125 | $8.6 \%$ | | 里具R\&R | 0.0136 | $9.3 \%$ | | 零件间 | 0.1451 | $99.6 \%$ |
  Concentrate:
Study the number of = 0.1457 = 0.1457 =0.1457quad=0.1457 \quad data ( ndc ) = 15 ( ndc ) = 15 (ndc)=15(\mathrm{ndc})=15 gradings that vary from variance
   Result:
*Based on the NDO analysis of the differential classification number, it is judged that the measurement system has reliable resolution.
*Given that the sampled samples can be representative of process variation, the above calculations are based on the total variation of the study.
  Criteria for Judgment   conclusion
R & R % < 10 % R & R % < 10 % R&R% < 10%\mathrm{R} \& \mathrm{R} \%<10 \%   The Interior Measurement System is acceptable. ***\star
10 % < R & R % < 30 % 10 % < R & R % < 30 % 10% < R&R% < 30%10 \%<\mathrm{R} \& \mathrm{R} \%<30 \%
Depending on the importance of the application/the cost of the device/maintenance element, it may be acceptable.
基于应用的重要性/则量装置成本/维修素,可能是可接受的。| 基于应用的重要性/则量装置成本/维修素,可能是可接受的。 | | :--- |
N & R % > 30 % N & R % > 30 % N&R% > 30%\mathrm{N} \& \mathrm{R} \%>30 \%
Unacceptable, all efforts should be made to improve the child-testing system.
不可接受的一一应该作出各种努力来改 进测童系统。| 不可接受的一一应该作出各种努力来改 | | :--- | | 进测童系统。 |
判断准则 结论 R&R% < 10% 测里系统是可接受的。 *** 10% < R&R% < 30% "基于应用的重要性/则量装置成本/维修素,可能是可接受的。" N&R% > 30% "不可接受的一一应该作出各种努力来改 进测童系统。" | 判断准则 | | 结论 | | :---: | :--- | :---: | | $\mathrm{R} \& \mathrm{R} \%<10 \%$ | 测里系统是可接受的。 | $\star$ | | $10 \%<\mathrm{R} \& \mathrm{R} \%<30 \%$ | 基于应用的重要性/则量装置成本/维修素,可能是可接受的。 | | | $\mathrm{N} \& \mathrm{R} \%>30 \%$ | 不可接受的一一应该作出各种努力来改 <br> 进测童系统。 | |
  Follow-up: