BP Pump 2 Non-Invasive Blood Pressure Simulator and Tester BP Pump 2 无创血压模拟器和测试仪
Technical Data 技术数据
The BP Pump 2 provides dynamic blood-pressure simulations for testing adult and neonatal noninvasive blood pressure monitors, including both arm- and wrist-cuff types. BP Pump 2 提供动态血压模拟,用于测试成人和新生儿无创血压计,包括手臂和腕套类型。
The analyzer features a preset mode for simulation of most patient conditions and the capability to program user-defined simulations. BP Pump 2 tests for leaks, measures static pressure, generates pressure, and tests overpressure valves. For improved testing versatility, the analyzer’s recently upgraded waveform test suite includes additional physiological selections. 该分析仪具有用于模拟大多数患者状况的预设模式,并能够对用户定义的模拟进行编程。BP Pump 2 测试泄漏、测量静压、产生压力并测试超压阀。为了提高测试的多功能性,分析仪最近升级的波形测试套件包括额外的生理选择。
BP Pump 2 comes in two models: the standard BP Pump 2_(L)2_{L} and the BP Pump 2_(M)2_{M}, which features a high-accuracy pressure transducer. BP Pump 2 also includes an optional five-lead synchronized ECG simulations to test monitors that monitors patients ECG. BP Pump 2 有两种型号:标准 BP 泵 2_(L)2_{L} 和 BP 泵 2_(M)2_{M} ,后者具有高精度压力传感器。BP Pump 2 还包括一个可选的五导联同步心电图模拟,用于测试监测患者心电图的监视器。
Key features 主要特点
Dynamic simulations for arm- and wrist-cuff monitors 手臂和腕带监护仪的动态模拟
Physiological waveform 生理波形
Internal pump for use in high- and low-pressure release verification, leak testing and pressure sourcing 用于高压和低压释放验证、泄漏测试和压力来源的内部泵
Preset mode for simulation of most patient conditions 用于模拟大多数患者状况的预设模式
User-definable systolic and diastolic values, along with heart rate and pulse volume and user-defined autosequences 用户可定义的收缩压和舒张压值,以及心率和脉搏量以及用户定义的自动排序
Five-leads synchronized ECG simulation 五导联同步心电图模拟
Internal cuff volume for basic device testing 用于基本设备测试的内部袖带体积
Respiratory artifacts, including spontaneous breathing and controlled ventilation 呼吸系统伪影,包括自主呼吸和受控通气
Arrhythmia simulations, including premature atrial contractions #1 and #2, atrial fibrillation, and PVCs 心律失常模拟,包括心房早搏 #1 和 #2、心房颤动和 PVC
PC based ansur test automation system to standardize testing protocol and documentation 基于 PC 的 ansur 测试自动化系统,用于标准化测试协议和文档
High-accuracy pressure transducer (Bp Pump2 _(M){ }_{\mathrm{M}} version only) 高精度压力传感器(仅限 Bp Pump2 _(M){ }_{\mathrm{M}} 版本)
Dynamic pressure simulation repeatability within 2 mmHg at maximal pulse size independent of device under test 在最大脉冲大小下,动态压力模拟可重复性在 2 mmHg 以内,与被测设备无关
Within 2 mmHg (at maximal pulse size independent of device under test) 2 mmHg 以内(在最大脉冲大小下,与被测设备无关)
Heart rate 心率
30 BPM to 250 BPM 30 BPM 至 250 BPM
Pulse volume 脉搏量
0.1 cc to 2.4 cc in increments of 0.1 cc 0.1 cc 至 2.4 cc,以 0.1 cc 为增量
Simulation parameters performance 仿真参数性能
Max pulse volume 最大脉冲量
2.4 cc 2.4 毫升
Max heart rate 最大心率
200 BPM at 2.4 cc pulse volume; 250 BPM at 1.2 cc pulse volume 2.4 cc 脉冲量时为 200 BPM;250 BPM 在 1.2 cc 脉冲量下
Internal neonatal cuff volume 新生儿袖带内部体积
20 cc 20毫升
Internal adult cuff volume (including NN volume) 成人袖带内部体积(包括 NN 体积)
310 cc 310 毫升
Heart rate setting accuracy 心率设置精度
+-1BPM\pm 1 \mathrm{BPM}
Simulation units 模拟单元
kPa and mmHg (user selectable) kPa 和 mmHg(用户可选)
Pressure leak test 压力泄漏测试
The pressure port is pressurized from 0 mmHg to 400 mmHg and keeps track of the pressure loss over time. Peak pressure and present pressure are displayed at all times; leak rate is displayed when it is available. 压力端口的压力范围为 0 mmHg 至 400 mmHg,并跟踪压力损失随时间的变化。始终显示峰值压力和当前压力;泄漏率可用时显示。
Autosequences 自动排序
Nine autosequences are provided for four tests and up to five simulations 为四次测试和多达五次模拟提供了九个自动序列
Pressure generation/measurement
Static-pressure range 50 mmHg to 400 mmHg ( 53 kPa )
Difference between target pressure and actual pressure - 5 mmHg
Internal leak rate < 2mmHg per minute with minimum volume of 300 cc
Four respiratory artifacts 3 spontaneous breathing; controlled ventilation
3 adult wrist-cuff simulations Normal, Hyper, Hypo
Pressure source Specified pressure generated from 50 mmHg to 400 mmHg in selectable increments of 1 mmHg
Pressure gauge Static pressure measured from 0 mmHg to 400 mmHg at the pressure port
Pressure relief rest Test for the NIBPM pressure relief valve ( 0 mmHg to 400 mmHg ) with display of peak pressure
Neonate internal cuff simulations Internal neonate cuff; four standard neonate pressures
Neonate simulations
Cuff #1 "Blood pressure: 35/15
Heart rate: 120 BPM
Pulse volume: 0.1 cc"
Cuff #2 "Blood pressure: 60/30
Heart rate: 120 BPM
Pulse volume: 0.1 cc"
Cuff #3 "Blood pressure: 80/50
Heart rate: 120 BPM
Pulse volume: 0.1 cc"
Cuff #4 "Blood pressure: 100/70
Heart rate: 120 BPM
Pulse volume: 0.1 cc"
Normal sinus rhythm and arrhythmias
BP and ECG Healthy heart, weak pulse, mild exercise strenuous exercise, obese subject, geriatric subject, tachycardia, bradycardia irregular pulse
BP and ECG Premature atrial contractions # 1, premature atrial contractions # 2, premature ventricular contractions, atrial fibrillation and PVCs
User-definable simulations User-definable systolic and diastolic values, along with heart rate and pulse volume
Ranges
Systolic pressure range 20 mmHG to 250 mmHG
Diastolic pressure range 10 mmHG to 200 mmHG
Dynamic NIBP simulation repeatability Within 2 mmHg (at maximal pulse size independent of device under test)
Heart rate 30 BPM to 250 BPM
Pulse volume 0.1 cc to 2.4 cc in increments of 0.1 cc
Simulation parameters performance
Max pulse volume 2.4 cc
Max heart rate 200 BPM at 2.4 cc pulse volume; 250 BPM at 1.2 cc pulse volume
Internal neonatal cuff volume 20 cc
Internal adult cuff volume (including NN volume) 310 cc
Heart rate setting accuracy +-1BPM
Simulation units kPa and mmHg (user selectable)
Pressure leak test The pressure port is pressurized from 0 mmHg to 400 mmHg and keeps track of the pressure loss over time. Peak pressure and present pressure are displayed at all times; leak rate is displayed when it is available.
Autosequences Nine autosequences are provided for four tests and up to five simulations| Pressure generation/measurement | |
| :--- | :--- |
| Static-pressure range | 50 mmHg to 400 mmHg ( 53 kPa ) |
| Difference between target pressure and actual pressure | - 5 mmHg |
| Internal leak rate | $<2 \mathrm{mmHg}$ per minute with minimum volume of 300 cc |
| Four respiratory artifacts | 3 spontaneous breathing; controlled ventilation |
| 3 adult wrist-cuff simulations | Normal, Hyper, Hypo |
| Pressure source | Specified pressure generated from 50 mmHg to 400 mmHg in selectable increments of 1 mmHg |
| Pressure gauge | Static pressure measured from 0 mmHg to 400 mmHg at the pressure port |
| Pressure relief rest | Test for the NIBPM pressure relief valve ( 0 mmHg to 400 mmHg ) with display of peak pressure |
| Neonate internal cuff simulations | Internal neonate cuff; four standard neonate pressures |
| Neonate simulations | |
| Cuff #1 | Blood pressure: 35/15 <br> Heart rate: 120 BPM <br> Pulse volume: 0.1 cc |
| Cuff #2 | Blood pressure: 60/30 <br> Heart rate: 120 BPM <br> Pulse volume: 0.1 cc |
| Cuff #3 | Blood pressure: 80/50 <br> Heart rate: 120 BPM <br> Pulse volume: 0.1 cc |
| Cuff #4 | Blood pressure: 100/70 <br> Heart rate: 120 BPM <br> Pulse volume: 0.1 cc |
| Normal sinus rhythm and arrhythmias | |
| BP and ECG | Healthy heart, weak pulse, mild exercise strenuous exercise, obese subject, geriatric subject, tachycardia, bradycardia irregular pulse |
| BP and ECG | Premature atrial contractions # 1, premature atrial contractions # 2, premature ventricular contractions, atrial fibrillation and PVCs |
| User-definable simulations | User-definable systolic and diastolic values, along with heart rate and pulse volume |
| Ranges | |
| Systolic pressure range | 20 mmHG to 250 mmHG |
| Diastolic pressure range | 10 mmHG to 200 mmHG |
| Dynamic NIBP simulation repeatability | Within 2 mmHg (at maximal pulse size independent of device under test) |
| Heart rate | 30 BPM to 250 BPM |
| Pulse volume | 0.1 cc to 2.4 cc in increments of 0.1 cc |
| Simulation parameters performance | |
| Max pulse volume | 2.4 cc |
| Max heart rate | 200 BPM at 2.4 cc pulse volume; 250 BPM at 1.2 cc pulse volume |
| Internal neonatal cuff volume | 20 cc |
| Internal adult cuff volume (including NN volume) | 310 cc |
| Heart rate setting accuracy | $\pm 1 \mathrm{BPM}$ |
| Simulation units | kPa and mmHg (user selectable) |
| Pressure leak test | The pressure port is pressurized from 0 mmHg to 400 mmHg and keeps track of the pressure loss over time. Peak pressure and present pressure are displayed at all times; leak rate is displayed when it is available. |
| Autosequences | Nine autosequences are provided for four tests and up to five simulations |
0 mmHg to 300 mmHg : +-0.5%\pm 0.5 \% of reading +-1mmHg\pm 1 \mathrm{mmHg}
301 mmHg to 400mmHg:+-2%400 \mathrm{mmHg}: \pm 2 \% of reading
0 mmHg to 300 mmHg : +-0.5% of reading +-1mmHg
301 mmHg to 400mmHg:+-2% of reading| 0 mmHg to 300 mmHg : $\pm 0.5 \%$ of reading $\pm 1 \mathrm{mmHg}$ |
| :--- |
| 301 mmHg to $400 \mathrm{mmHg}: \pm 2 \%$ of reading |
Accuracy, BP Pump 2_(m)2_{\mathrm{m}} (high-accuracy version) 精度,BP 泵 2_(m)2_{\mathrm{m}} (高精度版本)
+-0.7mmHg\pm 0.7 \mathrm{mmHg} ( 0.09 kPa ) throughout range +-0.7mmHg\pm 0.7 \mathrm{mmHg} ( 0.09 kPa ) 整个范围
Parallel port 并行端口
25-pin female connector, with D-subminiature style and pinouts conforming to IBM PC printer port (unidirectional), HP and ASCII printers 25 针母头连接器,具有 D 超小型样式和符合 IBM PC 打印机端口(单向)、HP 和 ASCII 打印机的引脚排列
Wristsimulations
Simulation #1 "Blood pressure 120/80 (93)
Heart rate: 80 BPM
Pulse volume: 0.5 cc"
Simulation #2 Blood pressure 160/100 (120) Heart rate: 80 BPM Pulse volume: 0.5 cc
Simulation #3 Blood pressure: 80/55 (63) Heart rate: 80 BPM Pulse volume: 0.5 cC
Temperature
Operating 15^(@)C to 40^(@)C(59^(@)F:} to {:104^(@)F)
Storage -20^(@)C to 65^(@)C(-4^(@)F:} to {:149^(@)F)
Relative humidity 90^(@) max
Display Bright, large 4-line x 40-character alphanumeric display with back lighting
Dimensions (WxDxH) 25.4cmxx25.4cmx12.7cm ( 10inxx10inxx5in )
Weight 3.4 kg ( 7.5 lb )| Wristsimulations | |
| :--- | :--- |
| Simulation #1 | Blood pressure 120/80 (93) <br> Heart rate: 80 BPM <br> Pulse volume: 0.5 cc |
| Simulation #2 | Blood pressure 160/100 (120) Heart rate: 80 BPM Pulse volume: 0.5 cc |
| Simulation #3 | Blood pressure: 80/55 (63) Heart rate: 80 BPM Pulse volume: 0.5 cC |
| Temperature | |
| Operating | $15^{\circ} \mathrm{C}$ to $40^{\circ} \mathrm{C}\left(59^{\circ} \mathrm{F}\right.$ to $\left.104^{\circ} \mathrm{F}\right)$ |
| Storage | $-20^{\circ} \mathrm{C}$ to $65^{\circ} \mathrm{C}\left(-4^{\circ} \mathrm{F}\right.$ to $\left.149^{\circ} \mathrm{F}\right)$ |
| Relative humidity | $90^{\circ}$ max |
| Display | Bright, large 4-line x 40-character alphanumeric display with back lighting |
| Dimensions (WxDxH) | $25.4 \mathrm{~cm} \times 25.4 \mathrm{~cm} \mathrm{x} 12.7 \mathrm{~cm}$ ( $10 \mathrm{in} \times 10 \mathrm{in} \times 5 \mathrm{in}$ ) |
| Weight | 3.4 kg ( 7.5 lb ) |
Ordering information 订购信息
Model 型
BP Pump 2 _(L){ }_{\mathrm{L}} Non-Invasive Blood Pressure Analyzer (standard pressure transducer) BPPUMP2L-US120V United States, 120 V BPPUMP2L-AUS250V Australia, 250 V BPPUMP2L-DEN250V Denmark, 250 V BPPUMP2L-SHK250V Shuko, 250 v BPPUMP2L-ISR250V Israel, 250 V BPPUMP2L-ITAL250V Italy, 250 V BPPUMP2L-IND250V India, 250 V BPPUMP2L-SWZ250V Switzerland, 250 V BPPUMP2L-UK250V United Kingdom, 250 V BPPUMP2L-BRAZ250V Brazil, 250 V
BP Pump 2_(M)\mathbf{2}_{\mathbf{M}} Non-Invasive Blood Pressure Analyzer (high-accuracy pressure transducer) BPPUMP2M-US120V United States, 120 V BPPUMP2M-AUS250V Australia, 250 V BPPUMP2M-DEN250V Denmark, 250 V BPPUMP2M-SHK250V Shuko, 250 V BPPUMP2M-ISR250V Israel, 250 V BPPUMP2M-ITAL250V Italy, 250 V BPPUMP2M-IND250V India, 250 V BPPUMP2M-SWZ250V Switzerland, 250 V BPPUMP2M-UK250V United Kingdom, 250 V BPPUMP2M-BRAZ250V Brazil, 250 V BPPM2M/ECG-NIM (Includes a BPPUMP2 _(M^(-)){ }_{\mathrm{M}^{-}} AUS250V, a ECG Adapter Block, 100 ml and 500 ml rigid aluminum chambers and a manual pressure pump (700PMP))
BPPM2M-NIM (Includes a BPPUMP2 _("M "){ }_{\text {M }}-AUS250V, 100 ml and 500 ml rigid aluminum chambers and a manual pressure pump (700PMP)) BP Pump 2 _(L){ }_{\mathrm{L}} 无创血压分析仪(标准压力传感器)BPPUMP2L-US120V 美国、120 V BPPUMP2L-AUS250V 澳大利亚、250 V BPPUMP2L-DEN250V 丹麦、250 V BPPUMP2L-SHK250V Shuko、250 V BPPUMP2L-ISR250V 以色列、250 V BPPUMP2L-ITAL250V 意大利、250 V BPPUMP2L-IND250V 印度、250 V BPPUMP2L-SWZ250V 瑞士、250 V BPPUMP2L-UK250V 英国、250 V BPPUMP2L-BRAZ250V 巴西、250 V
BP Pump 2_(M)\mathbf{2}_{\mathbf{M}} 无创血压分析仪(高精度压力传感器)BPPUMP2M-US120V 美国、120 V BPPUMP2M-AUS250V 澳大利亚、250 V BPPUMP2M-DEN250V 丹麦、250 V BPPUMP2M-SHK250V Shuko、250 V BPPUMP2M-ISR250V 以色列、250 V BPPUMP2M-ITAL250V 意大利、250 V BPPUMP2M-IND250V 印度、250 V BPPUMP2M-SWZ250V 瑞士、250 V BPPUMP2M-UK250V 英国、250 V BPPUMP2M-BRAZ250V 巴西、250 V BPPM2M/ECG-NIM(包括 BPPUMP2 _(M^(-)){ }_{\mathrm{M}^{-}} AUS250V、心电图适配器块、100 毫升和 500 毫升硬质铝室以及手动压力泵 (700PMP))
BPPM2M-NIM(包括一个 _("M "){ }_{\text {M }} BPPUMP2 -AUS250V、100 毫升和 500 毫升硬质铝室以及一个手动压力泵 (700PMP))
BP Pump 2 _(L){ }_{\mathrm{L}} Non-Invasive Blood Pressure Analyzer with Test Automation (standard pressure transducer) BP Pump 2 _(L){ }_{\mathrm{L}} 无创血压分析仪,带测试自动化功能(标准压力传感器)
TA-BPPMP2L-US United States, 120 V TA-BPPMP2L-US 美国, 120 V
TA-BPPMP2L-AUS Australia, 250 V TA-BPPMP2L-AUS 澳大利亚, 250 V
TA-BPPMP2L-DEN Denmark, 250 V TA-BPPMP2L-DEN 丹麦,250 V
TA-BPPMP2L-SHK Shuko, 250 v TA-BPPMP2L-SHK Shuko,250 伏
TA-BPPMP2L-ISR Israel, 250 V TA-BPPMP2L-ISR 以色列,250 V
TA-BPPMP2L-ITAL Italy, 250 V TA-BPPMP2L-ITAL 意大利,250 V
TA-BPPMP2L-IND India, 250 V TA-BPPMP2L-IND 印度,250 V
TA-BPPMP2L-SWZ Switzerland, 250 V TA-BPPMP2L-SWZ 瑞士, 250 V
TA-BPPMP2L-UK United Kingdom, 250 V TA-BPPMP2L-UK 英国, 250 V
TA-BPPMP2L-BRAZ Brazil, 250 V TA-BPPMP2L-BRAZ 巴西,250 V
BP Pump 2_(M)\mathbf{2}_{\mathbf{M}} Non-Invasive Blood Pressure BP 泵 2_(M)\mathbf{2}_{\mathbf{M}} 无创血压
Analyzer with Test Automation (high-accuracy pressure transducer) 带测试自动化的分析仪(高精度压力传感器)
TA-BPPMP2M-US United States, 120 V TA-BPPMP2M-US 美国, 120 V
TA-BPPMP2M-AUS Australia, 250 V TA-BPPMP2M-AUS 澳大利亚, 250 V
TA-BPPMP2M-DEN Denmark, 250 V TA-BPPMP2M-DEN 丹麦, 250 V
TA-BPPMP2M-SHK Shuko, 250 v TA-BPPMP2M-SHK Shuko,250 伏
TA-BPPMP2M-ISR Israel, 250 V TA-BPPMP2M-ISR 以色列,250 V
TA-BPPMP2M-ITAL Italy, 250 V TA-BPPMP2M-ITAL 意大利,250 V
TA-BPPMP2M-IND India, 250 V TA-BPPMP2M-IND 印度,250 V
TA-BPPMP2M-SWZ Switzerland, 250 V TA-BPPMP2M-SWZ 瑞士, 250 V
TA-BPPMP2M-UK United Kingdom, 250 V TA-BPPMP2M-UK 英国, 250 V
TA-BPPMP2M-BRAZ Brazil, 250 V TA-BPPMP2M-BRAZ 巴西,250 V
BP Pump 2 _(L){ }_{\mathrm{L}} Non-Invasive Blood Pressure Testing Bundles (with 5 leads ECG Block) 血压泵 2 个 _(L){ }_{\mathrm{L}} 无创血压测试包(带 5 导联心电图阻滞)
BPPM2L/ECG-US United States, 120 V BPPM2L/ECG-US 美国, 120 V
BPPM2L/ECG-AUS Australia, 250 V BPPM2L/ECG-AUS 澳大利亚, 250 V
BPPM2L/ECG-DEN Denmark, 250 V BPPM2L/ECG-DEN 丹麦, 250 V
BPPM2L/ECG-SHK Shuko, 250 v BPPM2L/ECG-SHK 舒科,250 V
BPPM2L/ECG-ISR Israel, 250 V BPPM2L/ECG-ISR 以色列,250 V
BPPM2L/ECG-ITAL Italy, 250 V BPPM2L/ECG-ITAL 意大利,250 V
BPPM2L/ECG-IND India, 250 V BPPM2L/ECG-IND 印度,250 V
BPPM2L/ECG-SWZ Switzerland, 250 V BPPM2L/ECG-SWZ 瑞士, 250 V
BPPM2L/ECG-UK United Kingdom, 250 V BPPM2L/ECG-UK 英国, 250 V
Quick NIBP monitor testing bundles 快速 NIBP 监测仪测试包
PS420/DPM1B Bundle PS420/DPM1B Bundle Kit (includes PS420, DPM1B, all accessories, and a custom carrying case) PS420/DPM1B 捆绑包 PS420/DPM1B 捆绑套件(包括 PS420、DPM1B、所有配件和定制手提箱)
PS410/DPM1B Bundle PS410/DPM1B Bundle Kit (includes PS410, DPM1B, all accessories, and a custom carrying case) PS410/DPM1B 捆绑包 PS410/DPM1B 捆绑套件(包括 PS410、DPM1B、所有配件和定制手提箱)
Standard accessories 标准配件
2780003FG Accessory Kit (tubings and fittings) 2780003FG 附件套件(管路和接头)
Operators Manual 作手册
Power Cord (country specific) 电源线(特定国家/地区)
Optional accessories 可选配件
ANSUR BP PUMP 2 Ansur BP Pump 2 Plug-in ANSUR BP PUMP 2 安苏尔 BP 泵 2 插件
5022010 Soft-Sided Vinyl Carrying Case 5022010 软边乙烯基手提箱
2780512FG ECG Adapter Block (allows simulation of 5-lead ECG waveforms) 2780512FG 心电图适配器模块(允许模拟 5 导联心电图波形)
71072 Parallel Printer Cable, D25M-C36M 71072 并行打印机电缆,D25M-C36M
PRINTR/414-US120V Printer, Seiko DPU-414-30B, 120 V power supply PRINTR/414-US120V 打印机,精工 DPU-414-30B,120 V 电源
PRINTR414-SHK220V Printer, Seiko PRINTR414-SHK220V 打印机,精工
DPU-414-30B, 220 V power supply DPU-414-30B,220 V 电源
61096 Printer, 120 V power supply 61096 打印机,120 V 电源
61097 Printer, 220 V power supply 61097 打印机,220 V 电源
97116 Printer Paper (7 rolls min) 97116 打印纸(至少 7 卷)
75034 Serial Cable, D9M-D9F 75034 串行电缆,D9M-D9F
5215-0269FG Adult Cuff Mandrel Spacer Block (three required) 5215-0269FG 成人袖带心轴垫片块(需要三个)
5215-0268FG Adult Cuff Mandrel End Block (two required) 5215-0268FG 成人袖带心轴端块(需要两个)
5027-0203FG Neonatal/External Cuff Mandrel (truncated plastic cylinder diameters: 7.6 cm , 10 cm , and 14 cm ) 5027-0203FG 新生儿/外袖带心轴(截头塑料圆柱体直径:7.6 cm、10 cm 和 14 cm)
98175FG Wrist Cuff Mandrel (adult) 98175FG 腕带心轴(成人)
BPPM2M-2001 500 ML Rigid Aluminum Chamber for NIBP Testing BPPM2M-2001 500 ML NIBP 测试用硬质铝室
BPPM2M-2002 100 ML Rigid Aluminum Chamber for NIBP Testing BPPM2M-2002 100 ML NIBP 测试用硬质铝室
700PMP External Pressure Pump 700PMP 外置压力泵
About Fluke Biomedical 关于福禄克生物医学
Fluke Biomedical is the world’s leading manufacturer of quality biomedical test and simulation products. In addition, Fluke Biomedical provides the latest medical imaging and oncology quality-assurance solutions for regulatory compliance. Highly credentialed and equipped with a NVLAP Lab Code 200566-6 accredited laboratory, Fluke Biomedical also offers the best in quality and customer service for all your equipment calibration needs. 福禄克生物医学是世界领先的优质生物医学测试和模拟产品制造商。此外,福禄克生物医学还提供最新的医学成像和肿瘤学质量保证解决方案,以实现法规遵从性。福禄克生物医疗拥有高度资质,并配备了 NVLAP 实验室代码 200566-6 认证的实验室,还为您的所有设备校准需求提供最佳的质量和客户服务。
Today, biomedical personnel must meet the increasing regulatory pressures, higher quality standards, and rapid technological growth, while performing their work faster and more efficiently than ever. Fluke Biomedical provides a diverse range of software and hardware tools to meet today’s challenges. 如今,生物医学人员必须满足日益增长的监管压力、更高的质量标准和快速的技术发展,同时比以往更快、更高效地开展工作。福禄克生物医学提供各种软件和硬件工具来应对当今的挑战。
As a medical test device manufacturer, we recognize and follow certain quality standards and certifications when developing our products. We are ISO 9001 and ISO 13485 medical device certified and our products are: 作为医疗测试设备制造商,我们在开发产品时认可并遵循某些质量标准和认证。我们通过了 ISO 9001 和 ISO 13485 医疗器械认证,我们的产品是:
CE Certified, where required CE 认证(如有需要)
NIST Traceable and Calibrated NIST 可追溯和校准
UL, CSA, ETL Certified, where required UL、CSA、ETL 认证(如有需要)
NRC Compliant, where required 符合 NRC 标准(如果需要)